Specifications

19
TRAVELING WITH THE SYSTEM
Changes in
Elevation
Elevation changes can affect the pressure being delivered
by the Solo. At elevations less than 2,500 ft. (762 m), the Solo
should be set to the low setting. In areas with elevations of 2,500 ft.
(762 m) or higher, the Solo should be set to the high setting.
To verify the elevation setting:
1. Make sure the Solo is plugged in and the air flow is turned OFF.
2. Press the ramp button once and release it. The green signal light will
flash either one time for the low elevation setting (<2,500 ft.) or three
times for the high elevation setting (2,500 ft.).
To change the elevation setting:
1. Make sure the Solo is plugged in and the air flow is turned OFF.
2. Press the ramp button and hold it down. While holding the
ramp button down, press and release the pressure on/off button.
Release the ramp button.
3. Verify the elevation setting. Press the ramp button once and release it.
The green signal light will flash to verify the elevation setting.
one flash - low elevation setting (<2,500 ft.)
three flashes - high elevation setting (2,500 ft.)
Ramp
Button