Specifications
3
Specifications
Dimensions: 10.00" x 5.50" x 4.75" (25x14x12 cm)
Weight: < 5 lb. (2.27 kg)
Operation Transport & Storage
Temperature: 41 - 104° F (5 - 40° C) -4 - 140° F (-20 - 60° C)
Humidity: 30 - 75% Non-condensing 15 - 95% Non-condensing
Atmospheric Pressure: 76.7 - 102 kPascals 50 - 102 kPascals
The electrical installation of the relevant room must comply with
IEC requirements.
100-120 VAC / 230-240 VAC, 50/60 Hz, 1.0 A max.
12 VDC, 3 A DC max.
Type of protection against electric shock: Class II Equipment
Degree of protection against electric shock: Type BF Applied Part
Degree of protection against harmful ingress of water: Ordinary
Equipment (IPXO)
Mode of Operation: Continuous
Device Fuses:
Line Voltage Fuse Type
115 VAC (100-120 VAC) T 800 mA, 250 V, 5 x 20 mm
230 VAC (230-240 VAC) T 315 mA, 250 V, 5 x 20 mm
DC Cord Fuse (not included): 3 A, Time Delay, Type MDL
EMC Interference: Meets CISPR II, Class B Emissions Requirements
Use only Respironics fuses supplied by your home care provider.
Device Size
Product Use,
Transport, and
Storage
Power
Requirements
IMPORTANT!
0123
Pressure ON / OFF
Ramp
DC Power
Type BF Applied Part
Fuse
Class II
(Double Insulated)
Attention, consult
accompanying documents
European CE Declaration of Conformity
!
Symbol Key