Specifications

24
TROUBLESHOOTING
Problem
The air flow from the
REMstar LX seems warm.
The REMstar LX is
making a faint humming
sound when the air flow is
turned off.
The noise level of the
REMstar LX has changed
to include unusual or
harsh sounds during
operation.
The REMstar LX will not
turn on.
The REMstar Plus LX has
hours recorded on the time
meter when it is delivered.
Solutions
Replace or clean the filter. Make sure the REMstar LX is away from
bedding or curtains that could block the flow of air around the REMstar
LX. Make sure the REMstar LX is away from heating equipment (e.g.,
forced air vents, radiators).
This is a normal characteristic of the REMstar LX. If the humming
sound is significantly louder than it usually is, contact your home care
provider or Respironics for directions on having the REMstar LX
serviced. Please have the REMstar LX’s serial number ready when
you call.
Contact your home care provider or Respironics for directions
on having the REMstar LX serviced. Please have the REMstar LX’s
serial number ready when you call.
Make sure that the REMstar LX is plugged into a working outlet.
Replace the REMstar LX’s fuses if the green signal light does not come
on. Verify that the correct voltage has been selected (i.e., if your
electrical line is 115 V you must use the 115 V voltage setting). Press
the Pressure On/Off button. If the new fuses blow out, contact your
home care provider. If you are using a DC battery, the battery may need
to be recharged or replaced or the fuses may need to be replaced.
The REMstar Plus LX is tested at the factory prior to shipment. It is
normal for up to 200 hours of testing time to be recorded on the
REMstar Plus LX.