ROUTINE MAINTENANCE
Changing
the Fuses
CAUTION!
If the REMstar LX does not turn ON when the power cord is plugged in,
the fuses may need to be replaced. The fuses are located in the
back of the REMstar LX above the power cord connector.
If you change the voltage setting, the fuses must be changed.
Unplug the REMstar LX before changing the fuses. Use only
Respironics fuses supplied by your home care provider.
1. Remove the fuse holder by releasing the small tabs on the sides of the
holder. You may need a small screwdriver to help remove the holder.
2. Replace the old fuses with new fuses of the same type and rating. Use
only Respironics fuses supplied by your home care provider. Always
replace both fuses.
Line Voltage: Fuse Type:
115 VAC (100-120 VAC) T 800 mA, 5 x 20 mm
230 VAC (230-240 VAC) T 315 mA, 5 x 20 mm
3. Insert the fuse holder back into the REMstar LX and snap into place.
WARNING!