Specifications

18
TRAVELING WITH THE SYSTEM
Changes in
Elevation
Elevation changes can affect the pressure being delivered by the
REMstar LX. At elevations less than 2,500 ft. (762 m), the REMstar LX
should be set to the low setting. In areas with elevations between 2,500 ft.
and 5,000 ft. the REMstar LX should be set to the medium setting. In areas
with elevations of 5,000 ft. or higher, the REMstar LX should be set to the
high setting.
To verify the elevation setting:
1. Make sure the REMstar LX is plugged in and the air flow is turned OFF.
2. Press the ramp button. The elevation setting will be displayed while the
ramp button is depressed.
LO - less and 2,500 ft.
M - 2,500 - 5,000 ft.
Hi - over 5,000 ft.
To change the elevation setting:
1. Make sure the REMstar LX is plugged in and the air flow is turned OFF.
2. Press the ramp button and hold it down. While holding the ramp button
down, press and release the pressure on/off button. Release the ramp
button.
3. Verify the elevation setting. Press the ramp button. The elevation setting
will be displayed while the ramp button is depressed.