Specifications
10
FIRST TIME SETUP
Power Cord
3. Connect the power cord.
Plug the socket end of the
power cord into the AC
power cord connector on the
back of the REMstar LX and
the pronged end of the power
cord into an electrical outlet.
IMPORTANT!
CAUTION!
An international plug adapter may also be required to make
your power cord compatible with the electrical outlet of the
country you are in. If further assistance is required, contact
Respironics International Customer Service department at
1-800-345-6443 (US and Canada) or 1-412-731-2100.
Do not use a voltage converter. It may cause damage to the
REMstar LX. Use only Respironics approved power cords. Inspect
the power cord often for any signs of damage. Replace a damaged
cord immediately.
The REMstar LX is activated when the power cord is connected. The
Pressure On/Off button only turns the blower on/off.
4. Verify the elevation setting. (The air flow must be turned off.)
Press the ramp button. The elevation setting will be displayed
while the ramp button is depressed.
LO = less than 2,500 ft.
M = 2,500 - 5,000 ft.
HI = 5,001 - 7,500 ft.
*over 7,500 ft. = Contact your home care provider to have
your pressure setting adjusted.
To change the elevation setting, press the ramp button and hold it
down. While holding the ramp button down, press and release the
pressure on/off button. Release the ramp button.
DO NOT use the REMstar LX if the display is erratic. Contact your
home care provider for further instructions.
CAUTION!
IMPORTANT!