GUIDANCE AND MANUFACTURERS DECLARATION - ELECTROMAGNETIC IMMUNITY
is device is intended for use in the electromagnetic environment specified below. e user of
this device should make sure it is used in such an environment.
IMMUNITY TEST IEC 60601 TEST
LEVEL
COMPLIANCE LEVEL ELECTROMAGNETIC
ENVIRONMENT -
GUIDANCE
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humid
-
ity should be at least
30%.
Electrical fast
Transient/burst
IEC 61000-4-4
±2 kV for power sup-
ply lines
±1 kV for input-out-
put lines
±2 kV for supply
mains
±1 kV for input/out-
put lines
Mains power quality
should be that of a
typical home or hos-
pital environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common mode
±1 kV differential
mode
±2 kV for common
mode
Mains power quality
should be that of a
typical home or hos-
pital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for
0.5 cycle
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5 sec
<5% U
T
(>95% dip in U
T
) for
0.5 cycle
40% U
T
(60% dip in U
T
) for 5
cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5 sec
Mains power qual-
ity should be that
of a typical home
or hospital environ-
ment. If the user of
the device requires
continued operation
during power mains
interruptions, it is rec-
ommended that the
device be powered
from an uninterrupt-
ible power supply or
a battery.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.