User manual

7-2 REMSTAR PLUS M SERIES USER MANUAL
FIGURE 7–1 REMOVING THE FILTERS
3. Examine the filter(s) for cleanliness and integrity.
4. Wash the gray foam filter in warm water with a mild detergent. Rinse thoroughly to remove
all detergent residue. Allow the filter to air dry completely before reinstalling it. If the foam
filter is torn, replace it. (Only Respironics-supplied filters should be used as replacement
filters.)
5. If the white ultra-fine filter is dirty or torn, replace it.
6. Reinstall the filters, inserting the white ultra-fine filter first if applicable, as shown in Figure
7–2.
Filter Area
Disposable Ultra-Fine
Filter (optional)
Reusable Gray
Foam Filter
(required)
FIGURE 7–2 REINSTALLING THE FILTERS
Caution: Never install a wet filter into the device. It is recommended that you
clean the filter in the morning and alternate using the two foam
filters provided with the system to ensure sufficient drying time for the
cleaned filter.