User manual

7-1REMSTAR PLUS M SERIES USER MANUAL
CHAPTER 7: CLEANING AND MAINTENANCE
is chapter describes how to clean the device and its filters and provides tips on traveling with
your REMstar Plus M Series system.
7.1 CLEANING THE DEVICE
Warning: To avoid electrical shock, always unplug the power cord from the wall
outlet or DC power source before cleaning the device.
Caution: Do not immerse the device in liquid or allow any liquid to enter the
enclosure, inlet filter, or any opening.
1. Unplug the device, and wipe the outside of the device with a cloth slightly dampened with
water and a mild detergent. Let the device dry completely before plugging in the power cord.
2. Inspect the device and all circuit parts for damage after cleaning. Replace any damaged parts.
7.2 CLEANING OR REPLACING THE FILTERS
Caution: Operating the device with a dirty filter may keep the system from working
properly and may damage the device.
Under normal usage, you should clean the gray foam filter at least once every two weeks and
replace it with a new one every six months. e white ultra-fine filter is disposable and should be
replaced after 30 nights of use or sooner if it appears dirty. DO NOT clean the ultra-fine filter.
Caution: Dirty inlet filters may cause high operating temperatures that may affect
device performance. Regularly examine the inlet filters as needed for integ
-
rity and cleanliness.
1. If the device is operating, stop the airflow by pressing the button. Disconnect the device
from the power source.
2. Remove the filter(s) from the enclosure by gently squeezing the filter in the center and pull-
ing it away from the device, as shown in Figure 7–1.