User manual

0 M Series Heated Humidier
probleM Why IT happened WhaT To do
The blue LED is
blinking.
. There is no communica-
tion between the therapy
device and the humidier.
2. The therapy device is
not seated properly on the
humidier platform.
. The heater plate is not
heating.
. Following the Daily Use section earlier in this
manual, check the outlet power and verify that
the humidier and therapy device are properly
plugged in. Make sure the AC power cord is
connected correctly to the power supply and
the power supply cord is correctly plugged
into the wall. If the problem continues to occur,
contact your home care provider. Return the
humidier, therapy device, and power supply
to your provider, so they can determine if the
problem is with the therapy device, humidier,
or power supply.
2. Remove the therapy device from the hu-
midier base and reconnect. Be sure to use the
guide on the humidier base to slide the therapy
device onto the base. Make sure the air inlet
port on the humidier connects securely to the
air outlet port on the device.
Make sure you removed the therapy device air
outlet port. It MUST be removed before con-
necting to the humidier.
To remove the port:
a. Put your thumb underneath the port and in-
sert your index nger into the port opening.
b. Pivoting from the bottom, unsnap the port
and pull it away from the device.
. Check the power connections to the humidi-
er. If the problem continues, contact your home
care provider or Respironics.
NOTE: If the humidier is functioning but the
water is not heating, you can temporarily use
the humidier as a passover humidier.
C
-
F
l
e
x
air outlet port