8-2 BIpap auto m SerIeS uSer manual
preSSure
Pressure Increments: 4.0 to 25.0 cm H
2
O (in 0.5 cm H
2
O increments)
Pressure Stability:
<10.0 cm H
2
O (±0.5 cm H
2
O)
≥10.0 cm H
2
O (±1.0 cm H
2
O)
Dynamic pressure accuracy is measured at the patient end of the circuit with a Whisper Swivel II
and varying flow conditions.
Measured in accordance with EN ISO 17510-1 @ 1/3, 2/3, and Pmax with BPM set to 10, 15,
and 20 BPM @ 20° C (±5° C), 50% RH (±5%), and an atmospheric pressure of 101.54 kPascals.
Maximum Flow: 35 LPM
Measured in accordance with EN ISO 17510-1 @ Pmax.
Control aCCuraCy
parameter range aCCuraCy
IPAP 4 to 25 cmH
2
O ± 0.5 cmH
2
O from static setpoint**
EPAP 4 to 25 cmH
2
O ± 0.5 cmH
2
O from static setpoint**
Ramp Duration 0 to 45 minutes ± 10% of setting
Rise Time 0 to 3* ± 25%
* Rise time settings are adjustable by the patient and the provider from 0 (off) to 3. Rise time
varies from 150 msec to 400msec dependent on patient setting and the pressure differential
from IPAP to EPAP. e rise time setting should be adjusted according to the patient comfort
level.
** Measured at the patient end of the circuit with a Whisper Swivel II exhalation device and no
patient flow.
dISpoSal
Dispose of the device in accordance with local regulations.