Chapter 7: CleanIng and maIntenanCe
is chapter describes how to clean the device and its filters and provides tips on traveling with
your BiPAP Auto M Series system.
7.1 CleanIng the devICe
Wa rning: To avoid electrical shock, always unplug the power cord from the wall
outlet or DC power source before cleaning the device.
Caution: Do not immerse the device in liquid or allow any liquid to enter the
enclosure, inlet lter, or any opening.
1. Unplug the device, and wipe the outside of the device with a cloth slightly dampened with
water and a mild detergent. Let the device dry completely before plugging in the power cord.
2. Inspect the device and all circuit parts for damage after cleaning. Replace any damaged parts.
7.2 CleanIng or replaCIng the fIlterS
Caution: Operating the device with a dirty lter may keep the system from working
properly and may damage the device.
Under normal usage, you should clean the gray foam filter at least once every two weeks and
replace it with a new one every six months. e white ultra-fine filter is disposable and should be
replaced after 30 nights of use or sooner if it appears dirty. DO NOT clean the ultra-fine filter.
Caution: Dirty inlet lters may cause high operating temperatures that may aect
device performance. Regularly examine the inlet lters as needed for
integrity and cleanliness.
1. If the device is operating, stop the airflow by pressing the button. Disconnect the device
from the power source.