User manual

5-3BIpap auto m SerIeS uSer manual
alert audIBle
IndICator
vISual
IndICator
devICe aCtIon poSSIBle CauSe patIent aCtIon
Patient
Reminder
None
Backlights:
Screen displays
a customized
reminder message
whenever the
device transitions
from the Active
state to the Standby
state.
The device continues
to operate.
Your home care
provider may
set a patient
reminder
scheduled to pop
up at a particular
time to remind
you to replace
your mask,
change your
lters, etc.
Press any of the display
buttons to acknowledge
the message and clear it. If
you do not acknowledge
the reminder, it will
disappear after 6 minutes,
and then reappear for three
consecutive days whenever
the device transitions from
the Active state to the
Standby state. If you still
do not acknowledge the
message, the reminder
period will reset and the
message will not reappear
until the next time the
Patient Reminder setting
expires.
5.2 trouBleShootIng
e table below lists some of the problems you may experience with your device or mask and possible
solutions to those problems.
proBlem Why It happened What to do
Nothing happens when you
apply power to the device.
The audible indicator does not
sound and the backlights on
the buttons do not light.
Theres no power at the
outlet or the device is
unplugged.
If you are using AC power, check the outlet power and verify that
the device is properly plugged in. Make sure the AC power cord is
connected correctly to the power supply and the power supply cord
is securely connected to the device’s power inlet. If the problem
continues to occur, contact your home care provider. Return both the
device and power supply to your provider, so they can determine if
the problem is with the device or power supply.
If you are using DC power, make sure your DC power cord and
battery adaptor cable connections are secure. Check your battery. It
may need recharged or replaced. If the problem persists, check the
DC cords fuse following the instructions supplied with your DC cord.
The fuse may need to be replaced. If the problem still occurs, contact
your home care provider.
If you are using a humidier, make sure you follow the instructions
for applying power that are provided with your humidier.