5-2 BIpap auto m SerIeS uSer manual
alert audIBle
IndICator
vISual
IndICator
devICe aCtIon poSSIBle CauSe patIent aCtIon
SmartCard
Error
Backlights:
Screen displays the
following ashing
error symbol:
The device continues
to operate and
provide therapy,
but data logging
is unavailable and
FOSQ is deactivated.
A problem
exists with the
SmartCard
inserted in the
accessory slot.
The card may
be improperly
inserted or the
data may be
corrupt.
Remove the SmartCard
to clear the alert. Conrm
that the card is properly
oriented, with the arrow
pointing towards the
device, as shown below,
and reinsert the SmartCard.
If the alert continues
to occur, remove the
SmartCard from the device
and contact your home
care provider. The card may
be corrupt and need to be
replaced.
Mask Leak
Backlights:
The words Mask
Leak ash on the
Active Display
screen.
The device continues
to operate.
The breathing
circuit is
disconnected or
there is a large air
leak.
Check your breathing
circuit connections and
reconnect the tubing if it
has come loose. Press any
of the display buttons to
clear the alarm. Make sure
your mask is on properly
before you press the
button to restart the
airow.
If the alert continues to
occur, disable the mask
alert setting following the
instructions in Chapter 4,
and continue your therapy
overnight. Contact your
home care provider the
next morning to have your
mask checked. You may
need a mask retting.
Auto O None The words Auto
Oash on the
Standby screen,
indicating that a
mask disconnect
has been detected.
The airow shuts
o and the device
enters the Standby
state approximately
45-60 seconds after
detection.
The mask has
been removed.
Put your mask back on
and press the button
to turn the airow on and
resume therapy.