4-8 BIpap auto m SerIeS uSer manual
e Patient Setup screens are described below.
– FOSQ
+
Setup
1. Standby Screen
The Standby screen appears when you rst apply power to the
device. You can access the Patient Setup menu from this screen by
pressing the + button. Once inside the Setup menu, press the
or buttons to cycle through the screens.
1 2 3
Bi-Flex Setting
1 2 3
Bi-Flex Setting
2. Bi-Flex Setting Screen
This screen displays the Bi-Flex level set by your home care
provider. This setting allows you to adjust the level of air pressure
relief that you feel when you exhale during therapy. Your home
care provider will either enable, disable, or lock Bi-Flex.
– If Bi-Flex is enabled and the setting is not comfortable for
you, you can increase or decrease the setting by pressing
the + or – buttons. The setting of 1 provides a small amount
of pressure relief, with higher numbers providing additional
relief. Press the + or – buttons to increase or decrease the
setting.
– If your provider has locked the Bi-Flex setting, you can view
the setting, but cannot change it. The second screen at left
shows Bi-Flex locked.
Note: This screen only displays if your home care provider has
enabled Bi-Flex.
off 1 2 3
Rise Comfort
3. Rise Time Comfort Setting Screen
Rise time is the time it takes for the device to change from
EPAP to IPAP. This screen allows you to adjust the rise time
so you can nd the desired setting. You can modify the Rise
Comfort Setting by 1 when either the + or – button is touched.
• 0 (o) reduces the Rise Time feature to the lowest setting (o
= 150 msec).
• 1 sets Rise Time to 1 (200 msec), and allows the patient to
change the setting.
• 2 sets Rise Time to 2 (300 msec), and allows the patient to
change the setting.
• 3 sets Rise Time to 3 (400 msec), and allows the patient to
change the setting.
This is only available if Bi-Flex has been disabled.
Note: This screen will not display if your provider has not
enabled Rise Time on your device.