3-3BIpap auto m SerIeS uSer manual
2. Connect the tubing to the mask:
a. If you are using a mask with a built-in exhalation port, connect the masks connector to
the flexible tubing, as shown in Figure 3–3.
Flexible
Tubing
Connector
Mask's
Connector
Exhalation Port
fIgure 3–3 ConneCtIng a maSK WIth BuIlt-In exhalatIon port
b. If you are using a mask with a separate exhalation device, connect the open end of the
flexible tubing to the exhalation device as shown in Figure 3–4. Position the exhalation
device so that the vented air is blowing away from your face. Connect the masks
connector to the exhalation device.
Exhalation
Device
Flexible
Tubing
Connector
fIgure 3–4 ConneCtIng a maSK WIth a Separate exhalatIon devICe
Wa rning: The exhalation device (e.g., Whisper Swivel II) or exhalation port (on
masks with an integrated exhalation port) is designed to exhaust
CO
2
from the patient circuit. Do not block or seal the ports on the
exhalation device.
Wa rning: If you are using a full face mask (i.e., a mask covering both your
mouth and your nose), the mask must be equipped with a safety
(entrainment) valve.
3. Attach the headgear to the mask. See the instructions that came with your headgear.