Specifications
13
EverGo User Manual
4. To set the device’s flow setting,
first touch the large number (such
as the 2 shown in Figure 9) in the
center of the screen.
5. Each time you press the flow
setting number, an up arrow
and/or down arrow
will appear on the screen.
(Examples are shown in Figures 16-17). To increase the flow
setting, press the up arrow. To decrease the flow setting, press the
down arrow.
e device can be set to: 1, 1½, 2, 2½, 3, 3½, 4, 4½, 5, 5½, and 6.
Warning: It is very important to set your device to your prescribed
level of oxygen flow. Do not increase or decrease your flow rate from
your prescribed level until you first consult with your physician.
Note: When it is turned on, the device will automatically start at
the flow rate setting used when the device was last turned off. As a
precaution, each time you start the unit, verify the flow setting.
6. Position the nasal cannula on your face, as shown below, and
breathe normally through your nose.
Figure 10. Correct position of cannula on your face.
Figure 9. Touch the number to change the
flow setting.