User manual
Provider Manual
46
Modes of Operation: Continuous
Electromagnetic Compatibility: The Harmony meets the
requirements of
EN 60601-1-2.
Fuses: There are no user-
replaceable fuses.
PRESSURE
Output: 4 to 30 cm H
2
O
CONTROL ACCURACY
Parameter Range Accuracy
IPAP 4 to 30 cm H
2
O ± 2.5 cm H
2
O*
EPAP 4 to 25 cm H
2
O ± 2.5 cm H
2
O*
CPAP 4 to 20 cm H
2
O ± 2.5 cm H
2
O*
Breath Rate 0 to 30 BPM Greater of ± 1
BPM or ± 10%
of the setting
Timed Inspiration 0.5 to 3.0 seconds ± 10% of the
setting
Ramp Duration 0 to 45 minutes ± 10% of the
setting
Rise Time 1 to 6** ± 25%*
* Measured at the patient end of circuit with a Whisper Swivel II
exhalation port and no patient flow. Dynamic pressure accuracy is ±
5 cm H
2
O measured at the patient end of the circuit with a Whisper
Swivel II and varying flow conditions.
** The range of values correspond to hundreds of milliseconds (e.g., a
setting of 4 indicates a Rise Time of .4 seconds).