Specifications
30
Functions
IPAP: Delivers set pressure continuously
EPAP: Delivers set pressure continuously
Modes: Spontaneous
Spontaneous/Timed
Timed
CPAP
Pressure and Flow Characteristics
Pressure Range: IPAP: 4 cm H
2
O to 20 cm H
2
O* at ± 60 L/min flow
EPAP: 4 cm H
2
O to 20 cm H
2
O* at ± 60 L/min flow
Pressure controls are coupled so that the unit will not deliver a higher EPAP than
IPAP level when the unit is operated in the S, S/T, or T modes.
Pressure Calibration Accuracy: ± 2 cm H
2
O of the indicated control setting
Pressure Setting Stability: ± 1 cm H
2
O over 8 hours
Note: For specifications, all pressure measurements were made at the blower outlet unless
stated otherwise.
*Measured at the end of a 6-foot section of smooth inner lumen tubing.
Cycling
Breaths per Minute (BPM): 4 to 30
Breathing Rate Accuracy: ± 10% or 2 BPM of the indicated reading, whichever is greater
% IPAP: 10 to 90%
% IPAP Setting Accuracy: ± 10% of the indicated setting
Minimum IPAP Time: Approximately 200 msec
S, S/T mode only
Maximum IPAP Time: 3 seconds ± 0.5 second at a constant flow rate condition of 100 L/min
S, S/T modes only
Minimum EPAP Time: 500 msec ± 75 msec
S, S/T modes only
Spontaneous Trigger Threshold: Less than 47.5 cc/sec above the average leak flow rate of 14 L/min
maintained for 30 msec
Specifications
Chapter 14 Specifications