8. a. Set the Function Selector Knob to the T mode.
b. Verify that the appropriate LED is illuminated.
c. Verify that the cycling is per control settings. Time the intervals to verify the function.
d. Verify that the BPM setting is 10 and change the % IPAP setting to 50.
e. Observe that the cycling time agrees with the control settings.
f. Change the BPM setting to 15.
g. Observe that the cycling times agree with the control settings.
If all of the parameters are within the stated operating specifications, the system is ready for
patient application.
Adjust the settings to the appropriate level for patient application prior to putting the circuit
on the patient.
Warning: If the BiPAP S/T System fails to perform within the stated specifications, it should be referred
for service.