Specifications
12
Warnings and Cautions
!
®
®
i
i
P
B
AP
®
Ventilatory Support System
115
230
Warnings and Cautions
!
®
®
i
i
P
B
AP
®
Ventilatory Support System
115
230
0000000h
Model 330000
Serial No.
000000
% IPAP Control
Active when the Function Selector Knob is positioned in the T mode only. This
control works in conjunction with the BPM control and determines the fraction
of the respiratory cycle spent in IPAP. Range = 10 - 90%
NOTE: For % IPAP values greater than 50%, time spent at IPAP will
exceed the time spent at EPAP (i.e., an inverse I:E ratio will be
delivered).
% IPAP
10 90
30
50
70
Description of Controls
Rear Panel
Control Panel Door
Contains the warnings, cautions, and electrical information
necessary for safe operation of the BiPAP S/T System.
When using the Detachable Control Panel, the door may be
left closed during operation.
Voltage Selector Switch
Set this switch to the appropriate line voltage.
U.S.A. and Canada - 115 Volts
European and other countries - Check the country’s line voltage
requirements for the correct setting.