Fuses: There are no user-replaceable fuses.
Type of Protection Against Electric Shock: Class II/Internally Powered Equipment
Degree of Protection Against Electric Shock: Type BF Applied Part
Degree of Protection against Ingress of Water
(Device and AC power supply): Drip Proof, IPX1
Mode of Operation: Continuous
Pressure
Pressure Increments: 4.0 to 25.0 cm H
2
O (in 1.0 cm H
2
O increments)
Flex Therapy Feature: O, 1, 2, 3
Control Accuracy
Parameter Range Accuracy
IPAP 4 – 25 cm H
2
O ± 2.5 cm H
2
O*
EPAP 4 – 25 cm H
2
O ± 2.5 cm H
2
O*
CPAP 4 – 20 cm H
2
O ± 2.5 cm H
2
O*
Breath rate 0 to 30 BPM greater of ± 1 BPM or ±10% of setting
Inspiration time 0.5 to 3
seconds
± (10% of setting + 0.1 second).
*Pressure measured at the patient connection port with or without the humidier (no patient ow,
with Whisper Swivel II).
Displayed Parameter Accuracy
Parameter Accuracy Resolution Range
Estimated Leak Rate ±(5+0.15 of reading) LPM 1 LPM 0 to 200 LPM
Exhaled Tidal Volume ±(25+0.15 of reading) ml 5 ml 0 to 2000 ml
Respiratory Rate Greater of ±1 BPM or
±10% of reading
1 BPM 0 to 60 BPM
Exhaled Minute Ventilation ±(1+0.15 of reading) LPM 1 LPM 0 to 99 LPM