Supplemental Oxygen
Oxygen may be added anywhere in the patient circuit provided that
a pressure valve is used. Please note the warnings listed below when
using oxygen with the device.
WARNINGS
When using oxygen with this system, the oxygen supply •
must comply with local regulations for medical oxygen.
When using oxygen with this system, turn the device •
on before turning on the oxygen. Turn the oxygen o
before turning the device o. This will prevent oxygen
accumulation in the device.
Explanation of the Warning: When the device is not
in operation and the oxygen ow is left on, oxygen
delivered into the tubing may accumulate within the
device’s enclosure. Oxygen accumulated in the device
enclosure will create a risk of re.
When using oxygen with this system, a Philips •
Respironics Pressure Valve (REF 302418) must be placed
at the air outlet port. Failure to use the pressure valve
could result in a re hazard. Refer to the pressure valve
instructions for use for proper use.
If administering xed-ow supplemental oxygen, the •
oxygen concentration may not be constant. The inspired
oxygen concentration will vary depending on the device
pressure settings, patient breathing pattern, and leak
rate. Substantial leaks around the mask may reduce the
inspired oxygen concentration to less than the expected
concentrations. Appropriate patient monitoring should
be implemented.
Oxygen supports combustion. Oxygen should not be •
used while smoking or in the presence of an open ame.
Do not connect the device to an unregulated or high •
pressure oxygen source.
Do not use the device near a source of toxic or •
harmful vapors.