User manual

BiPAP S/T user manual
36
Problem Why It Happened What To Do
The device display is
erratic.
The device has been
dropped or mishandled,
or the device is in
an area with high
Electromagnetic
Interference (EMI)
emissions.
Unplug the device. Reapply power to
the device. If the problem continues,
relocate the device to an area with lower
EMI emissions (away from electonic
equipment such as cellular phones,
cordless phones, computers, TVs,
electronic games, hair dryers, etc.). If the
problem still occurs, contact your home
care provider for assistance.
The device will not turn
o.
The correct blower
o sequence was not
followed.
Select Therapy to go back to the Monitor
Pressure screen. Push and hold the knob
for 2 seconds.
The Ramp feature does
not work when you press
the Ramp button.
Your home care provider
did not prescribe Ramp
for you, or your pressure
is already set to the
minimum setting.
If Ramp has not been prescribed for you,
the Ramp feature will not work.
If your provider has enabled Ramp but
the feature still does not work, check the
pressure setting on your Monitor Pressure
screen. If the pressure is set to the
minimum setting, or the starting pressure
is the same as the prescribed pressure,
the Ramp feature will not work.
The airow is much
warmer than usual.
The air lters may be
dirty.
The device may be
operating in direct
sunlight or near a heater.
Clean or replace the air lters.
The temperature of the air may vary
somewhat based on your room
temperature. Make sure the device is
properly ventilated. Keep it away from
bedding or curtains that could block the
ow of air around the device. Make sure
the device is away from direct sunlight
and heating equipment.
If using the humidier with the device,
check the humidier settings. Refer to the
humidier instructions to make sure the
humidier is working properly.