User manual

BiPAP S/T user manual
34
Alarm Priority Audible
Indicator
Visual Indicators Device
Action
User Action
SD Card
Removed
Info
SD card removed
- or - (if Icon option is
selected)
Operates This message occurs when
the SD card is removed from
the device. It is present for
30 seconds or until the user
acknowledges it. No action is
needed.
Humidier
Failure
Info None Flashing humidier LED
icon
Device
operate;
Humidier
shuts down
Alert is present for 12 minutes
or until the condition is xed.
Turn o airow and reconnect
the humidier to the device
according to the humidier
instructions. If the alert
continues to occur, contact
your home care provider.