User manual
BiPAP S/T user manual
34
Alarm Priority Audible
Indicator
Visual Indicators Device
Action
User Action
SD Card
Removed
Info
•
SD card removed
- or - (if Icon option is
selected)
Operates This message occurs when
the SD card is removed from
the device. It is present for
30 seconds or until the user
acknowledges it. No action is
needed.
Humidier
Failure
Info None Flashing humidier LED
icon
Device
operate;
Humidier
shuts down
Alert is present for 12 minutes
or until the condition is xed.
Turn o airow and reconnect
the humidier to the device
according to the humidier
instructions. If the alert
continues to occur, contact
your home care provider.