User manual
BiPAP S/T user manual
32
Alarm Priority Audible
Indicator
Visual Indicators Device
Action
User Action
Patient
Disconnect
High
• • • • •
Red ashing button
Patient Disconnect
- or - (if Icon option is
selected)
Operates This alarm occurs when the
patient circuit is disconnected
or has a large leak. Press
the Alarm Silence/Indicator
button to silence the alarm.
Reconnect the patient circuit
or x the leak. If the alarm
continues, contact your home
care provider for service.
Low Input
Voltage
Medium
• • •
Yellow ashing button
Low Voltage
- or - (if Icon option is
selected)
Ventilator
Operates;
Humidier
shuts down
The alarm is caused when
input power at the device,
either from an AC outlet
or battery, falls below the
acceptable limit for 10
seconds.
Press the Alarm Silence/
Indicator button to silence the
alarm. If the device is plugged
into a wall outlet, unplug the
device and then plug it back
in. If the alarm continues to
occur, contact your home care
provider for service.
If you are using a battery,
replace the battery or plug the
device into an AC outlet. If the
alarm continues, contact your
home care provider for service.
SD Card
Corrupted
Low
• •
Solid yellow button
SD card corrupted
- or - (if Icon option is
selected)
Operates This alarm occurs when a
problem exists with the
SD card. The data may be
corrupted. Press the Alarm
Silence/Indicator button to
silence the alarm. Contact your
home care provider with any
questions.