Alarm Message Screens
When an alarm message is activated, an alarm screen is displayed,
showing the text or icon specic to the most recent, highest priority
alarm.
Pressing the Control Wheel will reset the alarm and remove the alarm
screen from the display. Resetting the alarm allows you to return to
the previous screen. If multiple alarms occur during the same period
of time, the alarm screen will display the higher priority alarm (higher
priority alarms take precedence over lower priority alarms).
Alarm Summary Table
The following table summarizes all of the high, medium, and low
priority alarms and informational messages.
Alarm Priority Audible
Indicator
Visual Indicators Device
Action
User Action
Loss of
Power
High
• •
Red ashing button; Blank
screen
Shuts down Remove your mask. Check
your power connections.
Make sure there is power at
the outlet or power source.
Restore power to the device. If
the alarm continues, contact
your home care provider for
service.
Ventilator
Inoperative
High Red solid button;
Ventilator Inoperative
- or - (if Icon option is
selected)
Shuts down Remove your mask. Press the
Alarm Silence/Indicator button
to silence the alarm. Contact
your home care provider for
service.
Note: Pressing the Control
Wheel resets all alarms.
Note: If the alarm pop-up is
present, you cannot see the
Monitor Pressure screen.