Alarm LED Indicators
The Alarm Silence/Indicator button lights up as follows whenever an
alarm is detected:
Red Flashing Indicator – High priority alarm is detected. •
Yellow Flashing Indicator – Medium priority alarm is •
detected.
Yellow Solid Indicator – Low priority alarm is detected. •
The Alarm Silence/Indicator button does not light up when
informational messages display.
Alert Audible Indicators
An audible indicator sounds whenever a power failure or a high,
medium, or low priority alarm is detected. Additionally, an audible
indicator sounds for informational messages and to conrm that
certain actions have occurred (for example, when an SD card is
inserted or removed from the device).
Ventilator Inoperative – When a ventilator inoperative •
alarm occurs, a continuous audible indicator sounds.
The alarm descriptions later in this chapter display this
indicator as:
Power Failure – When a power failure occurs, a series of •
beeps sounds in a 1 beep pattern, repeating one second
on, then one second o. The alarm
descriptions later in
this chapter display this indicator as: • •
High Priority – When a high priority alarm is active, a •
series of beeps sounds in the following pattern, which is
repeated twice: 3 beeps, a pause, and then 2 more beeps.
This indicator continues until the cause of the alarm is
corrected or the audible alarm is silenced. The alarm
descriptions later in this chapter display this indicator as:
• • • • • • • • • •