User manual

BiPAP S/T user manual
6
Cautions
A caution indicates the possibility of damage to the device.
Electrostatic
Discharge (ESD)
Pins of connectors should not be touched. Connections should not
be made to these connectors unless ESD precautionary procedures
are used. Precautionary procedures include methods to prevent
build-up of electrostatic charge (e.g., air conditioning, humidication,
conductive oor coverings, non-synthetic clothing), discharging ones
body to the frame of the equipment or system or to earth or a large
metal object, and bonding oneself by means of a wrist strap to the
equipment or system or to earth.
Before operating the device, ensure that the SD card cover is replaced
whenever any of the accessories such as the Link Module or modem
are not installed. Refer to the instructions that came with your
accessory.
Do not use antistatic or conductive hoses or conductive patient tubing
with the device.
Condensation Condensation may damage the device. If the device has been exposed
to either very hot or very cold temperatures, allow it to adjust to room
temperature (operating temperature) before starting therapy.
Extension Cords Do not use extension cords with this device.
Device Placement Do not place the device in or on any container that can collect or hold water.
Do not place the device directly onto carpet, fabric, or other ammable
materials.
Do not plug the device into an outlet controlled by a wall switch.
Air Filter A properly installed, undamaged reusable foam inlet lter is required
for proper operation.
A dirty inlet lter may cause high operating temperatures that may
aect device performance. Regularly examine the inlet lter as needed
for integrity and cleanliness.
Never install a wet lter into the device. You must ensure sucient
drying time for the cleaned lter.
Cleaning Do not immerse the device in liquid or allow any liquid to enter the
enclosure or inlet lter.