User manual
46
User Manual
pr e s s u r e
Output: 4 to 30 cm H
2
O
Co n T r o l aC C u r a C y
Parameter Range Accuracy
IPAP 4 to 30 cm H
2
O* ±5 cm H
2
O**
EPAP 4 to 25 cm H
2
O* ±5 cm H
2
O**
CPAP 4 to 20 cm H
2
O ±5 cm H
2
O**
Breath Rate 0 to 30 BPM Greater of ± 1 BPM or
±10% of the setting
(when measured over a
4 minute period)
Timed Inspiration 0.5 to 3.0 seconds ±(0.1 + 10% of the
setting) seconds
Ramp Duration 0 to 45 minutes ±10% of the setting
Rise Time 1 to 6*** ±25%****
* Limited to 20 cm H
2
OwhenusingtheBi-FlexfeatureinSmode.
** Dynamic pressure accuracy is ± 5 cm H
2
O measured at the patient end
ofthecircuitwithaWhisperSwivelIIandvaryingowconditions.
Static pressure accuracy is ± 2 cm H
2
O measured at the patient end of
thecircuitwithaWhisperSwivelIIandnopatientow.
*** The range of values correspond to tenths of seconds (e.g., a setting of
4 indicates a Rise Time of 0.4 seconds).
****MeasuredatthepatientendofcircuitwithaWhisperSwivelII
exhalationdeviceandnopatientow.