Explanation of the Warning: e device is intended to be used with special masks or connectors
that have exhalation ports to allow continuous fl ow of air out of the mask. When the device is
turned on and functioning properly, new air from the device fl ushes the exhaled air out through
the mask exhalation port. However, when the device is not operating, enough fresh air will not
be provided through the mask, and exhaled air may be rebreathed. Rebreathing of exhaled air for
longer than several minutes can in some circumstances lead to suff ocation.
When the device is used with an external humidifi er, position the humidifi er so that the water level
in the humidifi er is lower than you, and the humidifi er is on the same level or lower than the device.
Do not open the device enclosure.  ere are no user serviceable parts inside. Repairs and internal
servicing should only be performed by an authorized service agent.
Periodically inspect electrical cords for damage or signs of wear.
To avoid electrical shock, unplug the device before cleaning.
If you detect any unexplained changes in the performance of the device, if it is making unusual
or harsh sounds, if it has been dropped or mishandled, if the enclosure is broken, or if water has
entered the unit, discontinue use and contact your home care provider.
Pins of connectors identifi ed with the ESD warning symbol should not be touched. Connections
should not be made to these connectors unless ESD precautionary procedures are used.
Precautionary procedures include methods to prevent build-up of electrostatic discharge (e.g., air
conditioning, humidifi cation, conductive fl oor coverings, non-synthetic clothing), discharging
ones body to the frame of the equipment or system or to earth or a large metal object, and
bonding oneself by means of a wrist strap to the equipment or system or to earth.
Do not use the device if the room temperature is above 95° F (35° C). If the device is used at
room temperatures above 95° F (35° C), the temperature of the airfl ow may exceed 106° F (41°
C), which could cause irritation to your airway.
Do not operate the device in direct sunlight or near a heating appliance because these conditions
can increase the temperature of the air coming out of the device.
In order to ensure proper protection against electrical shock, only communications accessories
with an IEC 60601-1 approved power supply may be connected through the SleepLink interface.
All IEC 950 devices must only be connected to the 7-pin connector with the Respironics
Isolation cable (Part Number 1012865).
Oxygen supports combustion. Oxygen should not be used while smoking or in the presence of
an open fl ame.
Do not use the device in the presence of a fl ammable anaesthetic mixture in combination with
oxygen or air, or in the presence of nitrous oxide.
If oxygen is used with the device, the oxygen fl ow must be turned off when the device is not in use.
Explanation of the Warning: When the device is not in operation and the oxygen fl ow is left
on, oxygen delivered into the tubing may accumulate within the devices enclosure. Oxygen
accumulated in the device enclosure will create a risk of fi re.
If you are using oxygen, the device must be equipped with the Respironics Pressure Valve. Failure
to use the Pressure Valve could result in a fi re hazard.