Service manual
PART VI HOW TO REPLACE THE PARTS
6-14
ENGLISH
6.10 REPLACE THE POWER CORD
When you need to change the electrical power cord, you can proceed with
the following steps.
- Stop the operation and unplug the power plug from the electrical outlet.
- Open the bottom plate.
- Unplug white Live line and black Neutral wire connections from the
interlock switch wire connections.
- Unscrew the screw on the ground wire connection.
- Unscrew the screw on the power cord.
- Remove the strain relief bushing (power cord holder).
- Pull out the power cord wires.
1. Lay down the unit horizontally
with unit bottom led facing toward
you. Unscrew the 10 screws
located on the bottom of the unit
with Phillips screwdriver.
4. Open the screw on the
ground wires.
6. Grip the power cord holder with needle nose pliers and pull
out the power cord at the same time.
7. Pull out the power cord.
5. Unscrew the screw on the
power cord wires. Note: The
power cord wire is under the
motor ground wire.
3. Unplug the white wire (Live
line) connection from the
interlock switch.
2. Unplug the black wire
(Neutral wire) connection from
the interlock switch.
Required tools
Needle-nose pliers
Phillips
screwdriver
When you replace the electrical
power cord, simply do the
reverse operation.