Specifications
Section 3
3
Safety Warnings
Research Instruments Ltd
3
SECTION 3 - SAFETY WARNINGS
DO NOT disassemble or modify any part of the RI Witness Sperm Preparation Reader,
or substitute any component for any other. Doing so may result in damage to samples.
This voids the warranty and/or service contract.
WARNING To avoid the risk of electric shock, this equipment must only be connected
to a supply mains with protective earth.
ONLY use the power cable and power supply adaptor supplied with the system.The
cable to the power supply is the ‘disconnect device’ for this equipment. To remove
all electrical power from this product, disconnect the power cable from the electrical
outlet. Equipment should be positioned so as to allow easy access to the power
cable. The appliance coupler or mains plug is used as the disconnect and must
remain readily operable.
WARNING Not to be used in a patient environment.
The system should be operated by qualied and trained personnel only.
This symbol indicates cautionary text which should be followed to avoid injury to
users or damage to samples.
WARNING Use of this equipment adjacent to or stacked with other equipment
should be avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to verify
that they are operating normally.
WARNING There are no replaceable parts supplied with this device. Should any
parts need to be replaced, contact RI or your distributor.
WARNING Use of accessories, transducers and cables other than those specied
or provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
WARNING Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm (12
inches) to any part of the Sperm Preparation Reader, including cables specied by
the manufacturer. Otherwise, degradation of the performance of this equipment
could result.