User`s manual
31
Section 4 Diagnostic Test Procedure
This section describes the diagnostic test procedure using the E6000H diagnostic program.
4.1 System Set-Up for Diagnostic Program Execution
To execute the diagnostic program, use the following hardware; do not connect the user system interface board
and user system.
• E6000H (HS1544EPH60H)
• Host computer
• The E6000 PC interface board which will be one of the following boards:
PCI bus interface board (HS6000EIC01H or HS6000EIC02H)
PC card interface (HS6000EIP01H)
LAN adapter (HS6000ELN01H)
USB adapter (HS6000EIU01H or HS6000EIU02H)
1. Install the E6000 PC interface board in the host computer and connect the supplied PC interface cable to the
board.
2. Connect the PC interface cable to the emulator.
3. Connect the supplied AC power cable to the emulator.
4. Boot the host computer up and enter the command input wait state of DOS (the DOS prompt) (Windows
®
98SE or Windows
®
Me) or the command prompt (Windows NT
®
, Windows
®
2000, or Windows
®
XP). If the
prompt window is not set up to be displayed over the whole screen, press the [Alt + Enter] key chord to
switch the mode. To return the prompt to a window, press the [Alt + Enter] key. This key chord works in the
same way regardless of the OS being used.
Note: If the display of the MS-DOS prompt does not switch from full-screen to windowed or vice-versa in
response to the [Alt + Enter] key chord, mark the [Alt+Enter] check box in [Windows shortcut keys] of
the [Misc] page on the [MS-DOS Prompt Properties] dialog box and click the [Apply] button to update
the settings.