Technical information

( 49 / 76 )
4.4 Self-check
(1) Self-check Procedure
To run the emulator self-check, do so as explained here below. While the self-check is in progress,
LEDs will change as shown in Figure 4.4.
(1) Set the switches in the emulation pod same as the factory-setting (see Figure 4.3).
(2) When the target system is connected, disconnect the target system.
(3) Within 2 seconds of activating power to the emulator, press the RESET switch on the emulator
front panel to switch the emulator to maintenance mode.
(4) Check the "SAFE" LED starts blinking and then press the RESET switch again.
(5) The self-check will start. If the normal result is displayed in about 40 seconds, the self-check
terminated normally.
Figure 4.3 Switch settings when running the self-check (factory-settings)
(2) If an Error is Detected in the Self-check
If the self-check does not result normally (system status error or target status error shown in Figure
4.4), check the following.
•Check the connection of the emulation pod and the PC4701.
Download the proper firmware.
•Check if the switches in this product are set same as the factory-settings (Figure 4.3).
IMPORTANT
Note on the Self-check:
If the self-check does not result normally (excluding target system errors), the
emulation pod may be damaged. Then contact your local distributor.