Specifications
Error messages and solutions
\text\da590\errormes 8.2 DA590 Instruction Manual
Error message Cause Solution
NO FORMFEED RECEIVED
No form feed character (0C
hex) was received after an
address was sent.
Check the program setup in
your software.
ERROR LCA !!
CALL SERVICE
Error in loading the LCA
module.
Call for service !
CHECKSUM ERROR !!
MAKE DEFAULT RESET
Error in buffered RAM Switch the printer off.
Perform a default reset. (see
Sec. 5).
CHECKSUM ERROR !!
MAKE COUNTER RESET
Error in the buffered RAM for
ink status counter.
With this error message you
may continue to
use the printer.
Reset the ink status counter
with the "HEAD RESET" menu
in programming mode.
(see Sec. 5).
TEST MACRO TOO LONG
User-defined test address is
too long.
Limit the size of the test
address to 3 Kbytes.
If you are working in block
rotation (= “Rev” in the
display), the test address may
not exceed 8 lines.
BUFFER OVERFLOW !!
Error in data transfer Check the connection of the
interface cable.
PROM ERROR !!
VERIFY U9
Incorrect or missing Eprom
U9.
Check the corresponding
Eprom U9.
PROM ERROR !!
VERIFY U43
Incorrect or missing Eprom
U43.
Check the corresponding
Eprom U43.
Note on maintenance:
Make sure that the feed rollers are always kept clean.