Operator's Manual

Chapter 14 Specifications and standards
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 244
Temperature Variation:
Test Standard: RTCA/DO-160E Section 5 Cat C
Rate of Variation: 2 °C per minute.
Humidity:
Test Standard: RTCA/DO-160E Section 6 Cat A
Transport/Storage: 15 to 95% RH Non-condensing (tested for 48 hours at 38-50
C)
Operating: 15 to 95% RH Non-condensing (tested at the end of the
storage cycle)
Operational Shocks & Crash Safety:
Test Standard: RTCA/DO-160E Section 7 Cat B
Operational Shock: Para 7.2 (6g for 11ms saw-tooth wave, repeated 3 times in all
axis).
Crash Safety: 20g in all directions (sustained).
Vibration:
Test Standard: MIL-810-F rotary, fixed wing (jet and turboprop profile)
Main case material: PC-ABS
Flame: UL94V-0
14.3.2 Rechargeable battery
Battery life At least 10¾ hours with display brightness at 60% (default),
ECG, SpO
2
, ETCO
2
, IP x 2, temp x 2 and NIBP every 15
minutes.
At least 11½ hours with display brightness at 30%, ECG,
SpO
2
, ETCO
2
, IP x 2, temp x 2 and NIBP every 15 minutes.
Up to 14 hours with battery saving mode activated (display at
default 60% brightness) – typically 12½ hours with the display
active 50% of the time.
Nominal voltage 7.4 V
Charging voltage 8.4 V ±1%
Nominal capacity 10.2 Ah, 75.5 Wh
Weight 0.42 kg nominal
Dimensions 152 x 42 x 62 mm max
Shelf life Approximately 75% remaining after 1 year (before the charge
indicator light turns to Amber)
Operating temperature -20°C to +60°C
Note
Battery shelf life and run times are based on a new, fully charged battery stored and used at
20 C. Run time is based on RDT’s model of typical device usage in an incident.