14.3 Miscellaneous features and specifications
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 243
14.3 Miscellaneous features and specifications
14.3.1 Invasive pressure USB module 01-2017 (optional)
Channels 2
Sensitivity 5 µV/V/mmHg
Bridge
180 minimum
Response 0-20 Hz (-3 dB)
Filters 50/100/150 Hz & 50-60 Hz notch
Range: -99 – 310 mmHg
Accuracy: ± 2% or ±2 digits including RDT’s adaptor cable but excluding
the transducer
Resolution: 1 mmHg
Warm up time 10 seconds
Ingress protection IP56
Size 108 mm x 63 mm x 36 mm
Weight 116g
EMC IEC 60601-1-2:2014 Class B
Altitude: -200 to 5486 m (-656 to 18000 ft)
(can be used at higher physical altitudes provided the local
atmosphere is no higher than 5486 m, e.g. in a pressurised
aircraft cabin)
Relative humidity: 15%-95% (non-condensing)
Operating temperature range: 0º C to 40º C due to transducer specifications
Transport/Storage temperature range: -37º C to +73.3º C
Transport/Storage pressure range: -200 to 5486 m (-656 to 18000 ft); 104 kPa to 53 kPa
Drop 1 m drop per IEC60601-1
Transit Drop 1.22 m (4 ft) MIL 810G, 26 drops all corners, faces and edges
with and without accessories connected (when not connected
to the Tempus Pro)
Impact 500 g steel ball from 1.3 m as per IEC60601
Emission of radio Frequency energy: RTCA/DO-160G Section 21 Cat M.
Temperature & Altitude:
Category: A1
Test Standard, Temperature, RTCA/DO-160E Section 4, Para 4.5.1 to 4.5.4
Test Standard, Altitude, RTCA/DO-160E Section 4, Para 4.6.1 and 4.6.2
Temperature: Operating: 0 °C to +40 °C
Storage: -37 °C to +73.3 °C
Rapid Decompression: 2438 m to 15545 m (8,000 ft to 51,000 ft) in 1 second