14.2 Physical characteristics and environmental specifications
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 241
Water proofness:
Not tested to RTCA/DO-160G Section 10
Commercial qualification:
Fluids susceptibility:
Not applicable.
Sand & dust:
Commercial qualification: IP66 (dust ingress under pressure) – whole device (excluding
internal printer, which is IP43)
Fungus resistance:
Not applicable.
Salt spray:
Not applicable.
Magnetic effect:
Not applicable.
Power input:
Not tested to RTCA/DO-160E Section 16.
Commercial qualification:
EN61000
-
3
-
2:2006
inc A1/A2:2009
Mains harmonics
EN61000-3-3:2008 Mains flicker
EN61000-4-11:2004 voltage dips and interruptions
Voltage spike:
Not tested to RTCA/DO-160G Section 17.
Commercial qualification: IEC61000-4-4:2012 Fast transient bursts
Audio frequency conducted susceptibility – power inputs:
Not tested to RTCA/DO-160G Section 18.
Commercial qualification: IEC61000-4-5:2006 Surges
Induced signal susceptibility:
Not tested to RTCA/DO-160G Section 19.
Commercial qualification: IEC61000-4-6:2013 Conducted RF field
Radio frequency susceptibility (radiated & conducted):
Not tested to RTCA/DO-160G Section 20.
Commercial qualification: IEC61000-4-3:2006 inc A2:2010 Radiated RF Interference
Emission of radio frequency energy:
Test Standard: RTCA/DO-160G Section 21 Cat M.