Operator's Manual
Table Of Contents
14.2 Physical characteristics and environmental specifications
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 239
14.2 Physical characteristics and environmental specifications
Size and weight – non-printer model: Standalone size and weight of Tempus Pro part number
00-1004-R with its battery and RapidPak clip but without
external cables, sensors or accessories:
263 mm (10.3”) x 216 mm (8.5”) high x 100 mm (3.9”)
(width x height x depth)
2.9 kg (6.4 Ib) nominal
The non-printer model with a Bluetooth headset is slightly
heavier (part number 00-1007-R).
Size and weight – printer model: Standalone size and weight of Tempus Pro part number
00-1026-R with its battery and printer but without external
cables, sensors or accessories:
263 mm (10.3 ”) x 216 mm (8.5 ”) x 102 mm (4 ”)
(width x height x depth)
3.2 kg (7 Ib) nominal
Printer models without invasive pressure are slightly lighter
(part number 00-1024-R).
Altitude: -200 to 5486 m (-656 to 18000 ft)
(can be used at higher physical altitudes provided the local
atmosphere is no higher than 5486 m, e.g. in a pressurised
aircraft cabin)
Note
The power supply’s altitude rating must be adhered to.
Do not use the power supply outside of its specifications.
Relative humidity: 15%-95% (non-condensing)
Operating temperature range: 0º C to 50º C all functions, 0º C to 40º C charging
Short term operating low -20 °C for 60 mins after being stored
at room temperature
Transport/Storage temperature range: -37º C to +73.3º C
Transport/Storage pressure range: -200 to 5486 m (-656 to 18000 ft)
104 kPa to 53 kPa
Start-up time in normal use Time for switching on or starting until the equipment is ready
for normal use: 42 seconds
Start-up time from storage
Time required for the equipment to warm from the minimum
storage temperature between uses until the equipment is ready
for intended use at 20°C: 180 minutes
Time required for the equipment to cool from the maximum
storage temperature between uses until the equipment is ready
for its intended use at 20°C: 180 minutes
IP66 (dust and water ingress under pressure) – whole device. According to IEC60529, “IP” means ingress
protection. The following two numerals refer to grades of ingress protection for solid objects (the first
numeral) and water (the second numeral). The first numeral “6” means the device is “dust-proof” against the
ingress of dust to a size <75 µm. The second numeral means the device is protected against the ingress of
100 l/m of water “jetting” at the device from 2.5 to 3 m.