Operator's Manual

14.1 Specifications
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 235
Initial inflation pressure: 160 mmHg adult, 140 mmHg paediatric, 90 mmHg neonate
Method: Oscillometric, diastolic values correspond to Phase 5 Korotkoff
sounds
14.1.6 Invasive pressure
Channels up to 4 (2 as standard)
Sensitivity 5 µV/V/mmHg
Bridge
180 minimum
Response 0-20 Hz (-3 dB)
Filters 50-60 Hz notch
Range: -99 – 310 mmHg
Accuracy: ± 2% or ±2 digits including RDT’s adaptor cable but excluding
the transducer
Resolution: 1 mmHg
Warm up time 10 seconds
14.1.7 Masimo pulse oximetry
Technology Masimo
®
SET
®
for use in motion and low-perfusion
applications
Pulse Range 25-239 bpm
Pulse accuracy Accuracy (all ages): no motion ≤3 digits, motion ≤5 digits
Pulse resolution 1 bpm
Pulse averaging
8 seconds (configurable)
SpO
2
Range 1-100%
SpO
2
accuracy Accuracy (adults/child): no motion or low perfusion ±2 digits
70-100%, motion ±3 digits 70-100%
Accuracy (neonate): motion, no motion and low perfusion ±3
digits 70-100%
SpO
2
resolution 1%
Data update rate 3 seconds
Type Functional saturation (test methods available upon request)
Wavelength range Red 660 nm, infra-red 905 nm
Perfusion Index range 0.02-20%
Response <1 second delay
Signal strength range 0-7 bars
Plethysmogram 1 – 100, auto-gain
Patient population Reusable soft walled probe – for use in clinical and pre-
hospital applications, on fingers and toes for children/adults
over 8 years old (other probe types are available from
Masimo®)