Operator's Manual

Chapter 14 Specifications and standards
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 234
14.1.4 ETCO
2
sensor
The Capnometer is automatically compensated for local atmospheric pressure.
CO
2
Units mmHg or kPa
EtCO
2
Range 0-150 mmHg
CO
2
Waveform Resolution 0.1 mmHg
EtCO
2
, Resolution 1 mmHg
CO
2
Accuracy 0-38 mmHg: ± 2 mmHg
39-150 mmHg: ± (5% of reading + 0.08% for every 1 mmHg
above 38 mmHg)
Note
For operation above 35C, add an additional ± 1 mmHg
or ± 2% (whichever is greater).
Respiration Rate Range 0-149 bpm
Respiration Rate Accuracy 0-70 bpm: ±1 bpm
71-120 bpm: ±2 bpm
121-149 bpm: ±3 bpm
Flow Rate 50 (42.5 ≤ flow ≤ 65) ml/min, flow measured by volume
Waveform Sampling 20 samples/s
Response Time (warmed up) 7 s (typical) for ETCO2, <13 seconds for respiration
(assuming a breath rate of 15 rpm)
Calibration Interval Annual performance verification may be performed if required
and calibration then performed if the system requires.
Calibration required after 4000 hours’ usage of the
Capnometer.
14.1.5 Non-invasive blood pressure
Range: Adult Systolic 40 - 260 mmHg
Paediatric Systolic 40-230 mmHg
Neonate Systolic 40-130 mmHg
Adult Diastolic 20 - 200 mmHg
Paediatric Diastolic 20-160 mmHg
Neonate Diastolic 20-100 mmHg
Adult MAP 26 - 220 mmHg
Paediatric MAP 26-183 mmHg
Neonate MAP 26-110 mmHg
Accuracy: ± 3 mmHg
Resolution: 1 mmHg
Pulse rate range: 30-220 bpm ± 2% or ±3 bpm (whichever is greater)
Maximum inflation: 300 mmHg (150 mmHg neonate)