14.1 Specifications
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 233
Note that during monitoring prior and post recording ECG
frequency response filters will be 0.5 Hz – 175 Hz.
Defibrillator protection Patient leads are isolated from system and operator, with 5kV
protection
Electrosurgery protection Protected
Common Mode Rejection 95 dB
Leads Off Indicators Connection status for each lead is shown on acquisition
screen
Permanent filters during recordings High Pass: 0.05 Hz 1st order
Low Pass: 175 Hz 1st order
Note that during monitoring prior and post recording ECG
frequency response filters will be 0.5 Hz – 175 Hz.
Baseline Wander: Baseline reset by adaptive zeroing
algorithm
Notch filter (Mains Noise Rejection) 50 Hz 4th order Butterworth,
49.1Hz - 50.9 Hz,
60 Hz 4th order Butterworth,
59.1 Hz
-
60.9 Hz
Low pass (Muscle Artifact Filter) 35 Hz 4th order
14.1.3 Impedance pneumography (respiration)
Range: 30-150 rpm
Accuracy ±2 rpm or ±2% whichever is greater
Excitation current
<20 µA
Excitation frequency 56 kHz
Measurement Leads available Lead I & Lead II