Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 230
If the Tempus is set to operate in Demonstration Mode, there will be a label “Demo” placed on the home
screen and on the 12 Lead ECG recording screen (where the 50/60Hz mains filter buttons would normally
be).
If the Tempus is put into a Training Communications Mode, this will be indicated by the word “Training” to
the left of the iAssist connection process instruction.
In either case, the Tempus can be put back into normal operational mode by returning to the Demonstration
and Training Menu and resetting Demo Mode to “Off”. The correct communications mode can be reset by
accessing the Communications Mode Menu from the Main Menu.
WARNING
If a Tempus is set into either Demonstration or Training mode, it will remain set in this
way after the power has been cycled and after a new patient has been admitted.
Demonstration and Training Menu
Pressing here
turns ALL the
medical and non-
medical readings
to demo mode –
all readings will
be simulated
Pressing here
brings up the
Training Modes
Menu. This
gives a range of
communication
modes for use in
training, in which
the Tempus
simulates a full
data and voice
connection but
does not require
any real
equipment to
connect to