Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 220
11.4 Weekly checks
RDT recommends that the following weekly checks are performed on the Tempus:
CAUTION
To prevent battery failure, perform this check every week when the Tempus Pro is in use.
Remove the battery from the Tempus Pro and inspect its contacts. If contacts are damaged, replace
the battery. If contacts are dirty, clean them - see "1.1.6 Cleaning the battery contacts".
Perform a test print.
11.5 Troubleshooting
Occasionally, problems may occur with the Tempus Pro. Operator error, sensor problems or a failure within
the Tempus Pro could cause these problems. In most instances, the Tempus Pro will display an error
message on the screen. This section describes the possible error messages and what they mean.
All of the error messages take the form of a window which appears in the middle of the screen.
The window contains the following text:
A title which identifies the sensor or system which is having trouble;
A description of the problem;
The effect that the error will have on the performance of the Tempus;
Which button to press to clear the error message off the screen.
CAUTION
In the event that the Tempus Pro displays an error that is not described within this manual
e.g. application errors, turn the Tempus Pro off and then on again. This should clear the
error and allow normal operation to resume. Do not continue to use the device if such an
error is displayed. If symptoms persist, please contact RDT.