Operator's Manual

Chapter 10 After using the Tempus Pro
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 208
Liquid soap
Wex-cide®
Windex®
Never use any other cleaning agent than those described above. Use of more aggressive agents such as
alcohols, acetone and other solvents is not permitted.
The Tempus Pro instruments must be cleaned during the re-packing process.
In the event of contamination of the Tempus from human fluids or other waste matter the product should be
thoroughly decontaminated and disinfected. In this event the handle should be removed and discarded and
replaced with a new part sourced from RDT.
CAUTION
If the Tempus requires cleaning, users are advised to ensure that stiff or sharp objects are
NOT used to remove detritus from underneath the louvered vents in the front and rear
panels of the device. Doing so could compromise the IP66 sealing of the product.
The front panel vent is located where shown in the image below. The rear panel vents are
located behind the accessory pockets.
10.2.2 Cleaning the display
The screen may be cleaned using a proprietary screen cleaning wipe of the type used for other LCD
screens. Under no circumstances should any abrasive substance be applied to the screen. It should be
noted that the anti-reflective (AR) coating of the touchscreen may be impaired if unsuitable materials are
used to clean the touchscreen. The table below lists different cleaning materials and their effect on the AR
coating after a 30-minute exposure under ambient conditions:
Chemicals
Result
NaOH 3% aq. solution
Unchanged
Na4OH 3% aq. solution
Unchanged
HCI 3% aq. solution
Unchanged
Kitchen cleaner 5% aq. solution
Unchanged
Glass cleaner
Unchanged
Position of Vent
Front panel
vent position