Operator's Manual
Table Of Contents
9.2 Display options
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 169
Note
The second waveform in the large ECG view defaults to display the impedance
pneumography (ECG respiration) if the Capnometer is not plugged in.
Note
When switching between the 4 waveform and the large ECG views, the ECG gain and wave
speed does not change except if the gain is set to 2 mm/mV on the 4 channel view in which
case the ECG will be displayed as 4 mm/mV. The user can subsequently adjust the gain
settings to optimise the available waveform height. For details on changing gain settings,
see “6.1.8 ECG settings”.
9.2.3 High contrast mode
The first button below the different home screen views allows the user to enable a high contrast display
mode on the results screen. Enabling this allows the Tempus to display vital signs data in a black on white
display for high daylight conditions (shown below). The high contrast display mode can be enabled
automatically by pressing and holding the button for 2 seconds. This allows the user to switch the high
contrast display on by pressing a single button.
Note
The high contrast display is available for all Home screen display modes.
12 Lead ECG View