Operator's Manual
Table Of Contents
9.1 Data input and output
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 165
9.1.3 Tempus to Tempus data handover
The ’Tempus to Tempus data handover’ screen allows you to send or receive patient data using a USB
memory device or Tempus to Tempus USB data cable (part number 01-2243). The output is a complete
patient data package for handover from one Tempus to another. This allows Tempus data to move with the
patient and for the next level care giver to see the history of injuries and treatment in the field.
CAUTION
Do not import patient records back onto the originating Tempus (directly or once-removed).
Exporting / importing is for handing over data from one Tempus to another only.
Prerequisites – ensure that you have the following:
Tempus Pro to send patient data;
Tempus Pro to receive patient data;
USB memory device or Tempus to Tempus USB data cable (part number 01-2243).
Handover via USB memory device
To export patient data:
1. Insert a USB memory device into the sending Tempus Pro. The Tempus automatically opens the
Data Input/Output Menu screen.
2. Press Tempus to Tempus data handover. The Tempus opens the ‘Tempus to Tempus data
handover’ screen:
3. If the PDF patient report is required by the recipient, set Include pdf in export to yes.
4. Press Export for Handover. If the USB device contains any existing Tempus patient data records,
you will be asked if you want to keep or overwrite them.
5. The ‘Tempus to Tempus data handover’ screen displays a blue progress bar with status messages
.
When the export progress bar displays “Finished”, remove the USB memory device and hand it to
the next level care giver.
To import patient data:
Tempus to Tempus data handover – USB memory device
Press here to
start the export
process
Press here to
start the import
process