Operator's Manual

Chapter 14 Specifications and standards
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 258
14.6.3 Recommended safety distances
Recommended Safety Distances between portable and mobile RF Telecommunication Devices and the
Tempus Pro (Tab. 206 according to IEC EN 60601-1-2).
The TEMPUS PRO is intended for use in an electromagnetic environment with controlled RF disturbances.
The user of the device can help to avoid electromagnetic disturbances by keeping the minimum distance
between portable and mobile telecommunication devices (transmitters) and the device - depending on the
output power of the telecommunication devices as described below.
Safety Distance Depending on the Frequency in m
Nominal power of the
transmitter (W)
150 kHz to 80 MHz
d = 1.2√P
80 MHz to 800 MHz
d = 1.2√P
800 MHz to 2.7 GHz
d = 2.3√P
0,01 0.12 0.12 0.23
0,1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100
12
12
23
For transmitters with a maximum nominal power not mentioned above: To detect the recommended
safety distance use the equitation in the corresponding column. P is the maximum nominal power of the
transmitter in watt (W) according to the specifications of the transmitter manufacturer.
NOTE 1: For 80 Hz and 800 MHz the higher frequency range applies.
NOTE 2: These guidelines may not be applicable for all cases. The propagation of electromagnetic
values is influenced by absorptions and reflections of buildings, objects and people.
14.7 Factory default settings
Setting Range Factory Default
Communications
mode
Various – customer specific Uppermost listed mode within the
customer specific communications
mode menu
Patient Neonate, Child, Adult Adult
NIBP mode Manual, Auto, Rapid Cuff Auto
NIBP automatic
measurement interval
2, 3, 5, 10, 15, 30 and 60 minutes Automatic - 3 minutes
QRS volume 0-100% in steps of 20 0%
ECG waveform 1 Lead I-III – 3-lead ECG cable
Lead I-III, aVL, aVR & aVF –
4-lead ECG cable
Lead I-III, aVL, aVR, aVF & V –
5-lead ECG cable
Lead I-III, aVL, aVR, aVF, V1-6 –
12-lead ECG cable
Lead II