Operator's Manual

14.6 EMC information
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 253
Standard
Title
EN 60601-2-37 Medical electrical equipment - Part 2-37: Particular requirements for the
basic safety and essential performance of ultrasonic medical diagnostic
and monitoring equipment
IEC 60601-2-25 Medical electrical equipment - Part 2-25: Particular requirements for the
basic safety and essential performance of electrocardiographs
IEC 60601-2-27 Medical electrical equipment - Part 2-27: Particular requirements for the
basic safety and essential performance of electrocardiographic
monitoring equipment
EN 80601-2-61 Medical electrical equipment -- Part 2-61: Particular requirements for
basic safety and essential performance of pulse oximeter equipment
EN 80601-2-55
Medical electrical equipment -- Part 2-55: Particular requirements for
the basic safety and essential performance of respiratory gas monitors
EN 62304 Medical device software -- Software life cycle processes
EN 62366
Medical devices -- Part 1: Application of usability engineering to
medical devices
14.6 EMC information
The following tables provide information required to be provided under IEC60601-1-2.
14.6.1 Cable length of the sensors and the accessories
RDT Part
Number
Cable
Length
(typ.)
Tested
Length
Ethernet cable 01-2025 2.1m 2.1m
SpO2 sensor 01-2014 2.4m 2.4m
3-lead ECG cable Various 2.4m 2.4m
4-lead ECG cable Various 2.4m 2.4m
5-lead ECG cable Various 2.4m 2.4m
12-lead ECG cable Various 2.4m 2.4m
IBP adaptor cable Various 2.4m 2.4m
Wired headset 01-1019 1.2m 1.2m
Mains Power supply 01-2049 0.45m 0.45m
Mains lead 01-2055 2m 2m
WARNING
The use of longer cable lengths may
cause an increased emission or a reduced
interference resistance. The use of other sensors or cables except the ones